Hello friends,

Please find below main points discussed in today's meeting :

1. NICE password for SM18 user has been changed last night. The new password is Test-2005

2. Training Rules (in case of De-Training/Noise in voltage signals) will be discussed by experts, but meanwhile we will continue the same way.

3. During Magnetic Measurements, make sure to use "Red coloured Fast Stop" button as mentioned in the procedure before rotating the shafts manually.

4. During IAP at warm, in case of errors pertaining to cryo PLC (T300_air etc), check with cryo operators if their system is in correct phase for that magnet.

5. For CQW0, Template has been changed by Stafiniak to account for low ambient temperature outside.

6. Blower fans are put on all SSS benches once again as we faced problems last week because of condensation on Orbital dipole connectors.

7. Problem with Polarity selector switch on E-cluster has been solved. The cause cited by Eq.Support is that it was forced by someone in the past. Pl make sure to handle it gently.

8. In case of a problem during powering correctors, try to change Inductance value in TEMA somewhere between 1 mH to 20 mH.

Apart from this, few things related to SSS To-Do-List :

1. HV test at warm BEF2 - WP14 is being removed as it is being done in Bldg. 904. (Except two special cases for SSS74 and SSS100, )

2. Resistance measurements BEF3 - CQW0 will also be removed after few days. But, till then we will continue performing this test.

Thanks. Have a good day.

Anand.